HunterLab
UltraScan PRO (US PRO) is a high-performance benchtop spectrophotometer built for advanced research and quality control. With D65 illumination calibrated into the UV region, extended UV-VIS-NIR wavelength range, and diffuse/8° geometry, it accurately measures whitening agents, camouflage materials, UV blockers, and transmission haze. Its versatility, automated optics, and optional 21 CFR Part 11 compliance make it one of the most powerful color measurement systems available.
Inquiry| Specification | Details |
|---|---|
| Measurement Principle | Dual-beam spectrophotometer / colorimeter |
| Geometry | Diffuse d/8° reflectance, d/8° total transmission, d/0° regular transmission |
| Spectrophotometer / Colorimeter | Two polychromators, each with 512-element diode array and high-resolution concave holographic grating |
| Sphere Diameter | 152 mm (6 in.) |
| Sphere Coating | Spectraflect™ sphere, Duraflect™ port plate and specular exclusion door |
| Reflectance Port Sizes (RSIN / RSEX) | LAV: 25 mm illuminated / 19 mm measured; MAV: 13 mm illuminated / 9 mm measured; SAV: 7 mm illuminated / 4 mm measured |
| Total Transmission Port Sizes (TTRAN) | LAV: 25 mm illuminated / 17.4 mm measured; MAV: 25 mm illuminated / 13.2 mm measured; SAV: 25 mm illuminated / 11.6 mm measured |
| Regular Transmission Port Sizes (RTRAN) | LAV / MAV / SAV: 17 mm illuminated / 17 mm measured |
| Lens Switching | Automatic for LAV / MAV / SAV |
| Specular Component | Automated Included (RSIN) or Excluded (RSEX) |
| Spectral Range | 350 nm - 1050 nm |
| Wavelength Resolution | < 2 nm |
| Effective Bandwidth | 5 nm equivalent triangular |
| Reporting Interval | 5 nm |
| Photometric Range | 0 to 150 % |
| Light Source | Pulsed Xenon lamps (3), calibrated and controlled in UV range |
| Automatic UV Control | 400 nm cutoff filter for UV control / exclusion; optional 420 nm cutoff filter |
| Transmission Modes | Total (TTRAN) and Regular (RTRAN) |
| Transmission Compartment | Open on 3 sides; 10.2 cm D x 35.6 cm W x 16.5 cm H |
| Standards Conformance | Reflectance: CIE 15:2004, ISO 7724/1, ASTM E1164, DIN 5033, JIS Z 8722 C / Transmittance: CIE 15:2004, ASTM E1164, JIS Z 8722 E,G / Haze: ASTM D1003 |
| Standards Traceability | NIST traceable per CIE Publication 44 and ASTM E259 |
| Specification | Details |
|---|---|
| Colorimetric Repeatability (20 readings) | < 0.03 ΔE* on white tile (LAV); < 0.07 ΔE* on blue denim tile (LAV) |
| Spectral Repeatability | Max 0.20 range between 435 nm and 695 nm |
| Inter-instrument Agreement | ΔE* < 0.09 Avg / ΔE* < 0.20 Max on BCRA II Tile Set |
| Specification | Details |
|---|---|
| Dimensions | H: 32.3 cm (12.7 in.) / W: 42.0 cm (16.5 in.) / D: 49.8 cm (19.6 in.) |
| Weight | 25.9 kg (57 lbs) |
| Power | 90 to 250 VAC, 50 to 60 Hz / 60 W passive, 120 W maximum |
| Interface | RS-232C serial, 19,200 baud, DB9 female terminal |
| Operating Environment | 4° to 38°C (40° to 100°F), 10% to 85% RH, noncondensing |
| Storage Environment | -21° to 66°C (-5° to 150°F), 10% to 90% RH, noncondensing |
| Standard Accessories | White tile, certificate of traceability, black light trap, zero plate, green diagnostic tile, wavelength filter, fluorescent standard, sample clamp, LAV/MAV/SAV apertures, RS-232 cable, USB-serial adapter, power cord, EasyMatch QC software, manual |

HunterLab
CFL2 is a compact 45°/0° spectrophotometer for fast, accurate color measurement across a wide range of samples.

HunterLab
Advanced spectrophotometer for color-critical and fluorescent sample measurement.

HunterLab
Compact spectrophotometer for reliable non-contact color measurement in lab and production environments.

HunterLab
Transmission color and haze measurement in a single benchtop instrument.

HunterLab
US VIS is a spectrophotometer for precise reflected, transmitted color and haze measurement.