HunterLab
UltraScan VIS (US VIS) provides exceptional precision for opaque, translucent, transparent, dark, and highly saturated samples. Using diffuse/8° geometry with automated specular control, it measures reflected and transmitted color as well as haze in compliance with CIE, ASTM, and USP standards. Its outstanding long-term stability ensures confidence that color differences reflect true product changes—not instrument variability.
Inquiry| Specification | Details |
|---|---|
| Measurement Principle | Dual-beam benchtop spectrophotometer / colorimeter |
| Geometry | Diffuse d/8° reflectance, d/8° total transmission, d/0° regular transmission |
| Spectrophotometer / Colorimeter | Two 256-element diode arrays with high-resolution concave holographic grating |
| Sphere Diameter | 152 mm (6 in.) |
| Sphere Coating | Spectraflect™ sphere, Duraflect™ port plate and specular exclusion door |
| Reflectance Port Sizes (RSIN / RSEX) | LAV: 25 mm illuminated / 19 mm measured; SAV: 9.5 mm illuminated / 6 mm measured |
| Total Transmission Port Sizes (TTRAN) | LAV: 25 mm illuminated / 17.4 mm measured; SAV: 25 mm illuminated / 10 mm measured |
| Regular Transmission Port Sizes (RTRAN) | LAV / SAV: 17 mm illuminated / 17 mm measured |
| Lens Switching | Automatic for LAV / SAV |
| Specular Component | Automated Included (RSIN) or Excluded (RSEX) |
| Spectral Range | 360 nm - 780 nm |
| Wavelength Resolution | < 2 nm |
| Effective Bandwidth | 10 nm equivalent triangular |
| Reporting Interval | 10 nm |
| Photometric Range | 0 to 150 % |
| Photometric Resolution | 0.003 % (0.01 % reported) |
| Light Source | Pulsed Xenon lamp filtered to approximate D65 daylight |
| Automatic UV Control | 400 nm cutoff filter for UV control / exclusion; optional 420 nm cutoff filter |
| Flashes per Measurement | 1 in LAV mode / 4 in SAV mode |
| Measurement Time | < 5 seconds |
| Transmission Modes | Total (TTRAN) and Regular (RTRAN) |
| Transmission Compartment | Large and open on 3 sides; 10.2 cm D x 35.6 cm W x 16.5 cm H |
| Standards Conformance | Reflectance: CIE 15:2004, ISO 7724/1, ASTM E1164, DIN 5033, JIS Z 8722 C / Transmittance: CIE 15:2004, ASTM E1164, JIS Z 8722 E,G / Haze: ASTM D1003 |
| Standards Traceability | NIST traceable per CIE Publication 44 and ASTM E259 |
| Specification | Details |
|---|---|
| Colorimetric Repeatability (20 readings) | < 0.03 ΔE* on white tile (LAV / SAV); < 0.05 ΔE* on blue denim tile (LAV / SAV) |
| Spectral Repeatability | Max 0.20 peak-to-peak between 435 nm and 695 nm |
| Inter-instrument Agreement | ΔE* < 0.15 Avg / ΔE* < 0.25 Max on BCRA II Tile Set |
| Specification | Details |
|---|---|
| Dimensions | H: 27.9 cm (11 in.) / W: 42.0 cm (16.5 in.) / D: 49.8 cm (19.6 in.) |
| Weight | 20.4 kg (45 lbs) |
| Power | 100 to 240 VAC, 47 to 63 Hz / 60 W passive, 120 W maximum |
| Interface | RS-232C serial, 19,200 baud, DB9 female terminal |
| Operating Environment | 10° to 40°C (50° to 104°F), 10% to 90% RH, noncondensing |
| Storage Environment | -21° to 66°C (-5° to 150°F), 10% to 90% RH, noncondensing |
| Standard Accessories | White tile, certificate of traceability, black light trap, zero plate, green diagnostic tile, wavelength filter, sample clamp, LAV/SAV apertures, RS-232 cable, USB-serial adapter, power cord, EasyMatch QC software, manual |

HunterLab
CFL2 is a compact 45°/0° spectrophotometer for fast, accurate color measurement across a wide range of samples.

HunterLab
Advanced spectrophotometer for color-critical and fluorescent sample measurement.